Hispeed Lx//I, Hispeed Fx/I, and Hispeed Dx/I Family of Systems
510(k) K980169
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK980169 | Date13 Apr 1998 | StatusActive |
Recalls
-
- Terminated Z-2224-2009
During internal testing, GE Healthcare has recently determined that the collimator used on our CT/e and HiSpeed X/i families of Computed Tomography systems does not comply with the specific requirements in United States Title 21 of the Code of Federal Regulations, and IEC 60601-1-3, related to leakage radiation from the diagnostic source assembly, which includes the collimator. The excess leak
-
- Terminated Z-0269-05
The patient tables of the affected CT systems are supported by a hydraulic cylinder and a cylinder shaft (i.e. the upper shaft) which connects the tabletop to the table frame. GE received a report of a HiSpeed LX/I system table falling abruptly due to the filure of the upper shaft. No injuries were reported.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JAK
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent