Hitachi Altaire
by Hitachi Medical Systems America, Inc. 2 variants
510(k) K002420
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK002420 | Date17 Oct 2000 | StatusActive |
Recalls
-
- Terminated Z-1542-2008
The firm's Altaire MR Scanner's Emergency Rundown Switch Unit (ERDU), that is used to shut down the device in an emergency, was found to be defective. An ERDU failure could delay or prevent patient access leading to possible patient death or serious injury.
-
- Terminated Z-1263-2007
Overheating; A portion (balun assembly) of the cable that connects the shoulder coil to the actual MRI unit may malfunction and result in overheating of the balun assembly. The overheating of the balun assembly could result in injury to the patient or operator if either came into contact with the assembly.
-
- Terminated Z-1104-06
Software anomaly-A software defect can cause the slice line indicator to become mis-positioned during Multi-Planar Reconstructions (MPR) with the potential for patient mis-diagnosis.
-
- Terminated Z-0910-06
The device has a software anomaly which causes the operating system to 'lock up' and data to become corrupted during service procedures.
-
- Terminated Z-1103-06
Circuit Failure-There is the potential risk of the Phase Loss Detector Safety Device (PHSEN PCB) malfunctioning in the event of a power loss. The circuit may fail to turn off the system main breaker when 2 of the 3 power phases are lost.
-
- Terminated Z-0888-05
Image reconstruction software problem that could potentially cause multiple patient images to be stored in one patient folder.
-
- Terminated Z-0797-05
The electrical cables that supply gradient power to the magnet can become loose over time and create a potential fire hazard.
-
- Terminated Z-0846-05
The absorber unit in the helium compressor which is located in equipment rooms, and is not accessed by patients or system operators, may fail under pressure, causing an explosion that damages the compressor.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent