Hitachi Altaire

by Hitachi Medical Systems America, Inc. 2 variants

510(k) K002420

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

8 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Abbreviated NumberK002420 Date17 Oct 2000ClearedStatusActive

Recalls

8 recalls · FDA

    • Terminated Z-1542-2008 FDA Unclassified (pre-2012) initiated 11 Feb 2008

    The firm's Altaire MR Scanner's Emergency Rundown Switch Unit (ERDU), that is used to shut down the device in an emergency, was found to be defective. An ERDU failure could delay or prevent patient access leading to possible patient death or serious injury.

    • Terminated Z-1263-2007 FDA Unclassified (pre-2012) initiated 30 Jul 2007

    Overheating; A portion (balun assembly) of the cable that connects the shoulder coil to the actual MRI unit may malfunction and result in overheating of the balun assembly. The overheating of the balun assembly could result in injury to the patient or operator if either came into contact with the assembly.

    • Terminated Z-1104-06 FDA Unclassified (pre-2012) initiated 28 Apr 2006

    Software anomaly-A software defect can cause the slice line indicator to become mis-positioned during Multi-Planar Reconstructions (MPR) with the potential for patient mis-diagnosis.

    • Terminated Z-0910-06 FDA Unclassified (pre-2012) initiated 21 Feb 2006

    The device has a software anomaly which causes the operating system to 'lock up' and data to become corrupted during service procedures.

    • Terminated Z-1103-06 FDA Unclassified (pre-2012) initiated 27 Jan 2006

    Circuit Failure-There is the potential risk of the Phase Loss Detector Safety Device (PHSEN PCB) malfunctioning in the event of a power loss. The circuit may fail to turn off the system main breaker when 2 of the 3 power phases are lost.

    • Terminated Z-0888-05 FDA Unclassified (pre-2012) initiated 30 Jun 2005

    Image reconstruction software problem that could potentially cause multiple patient images to be stored in one patient folder.

    • Terminated Z-0797-05 FDA Unclassified (pre-2012) initiated 11 Apr 2005

    The electrical cables that supply gradient power to the magnet can become loose over time and create a potential fire hazard.

    • Terminated Z-0846-05 FDA Unclassified (pre-2012) initiated 18 Nov 2004

    The absorber unit in the helium compressor which is located in equipment rooms, and is not accessed by patients or system operators, may fail under pressure, causing an explosion that damages the compressor.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LNH
Device Class
Class II
Regulation
21 CFR Part 892.1000
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Hitachi Medical Systems America, Inc.

See all devices from Hitachi Medical Systems America, Inc. →