Hitachi Altaire
by Hitachi Medical Systems America, Inc. 2 variants
510(k) K022390
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK022390 | Date11 Oct 2002 | StatusActive |
Recalls
-
- Terminated Z-0910-06
The device has a software anomaly which causes the operating system to 'lock up' and data to become corrupted during service procedures.
-
- Terminated Z-1103-06
Circuit Failure-There is the potential risk of the Phase Loss Detector Safety Device (PHSEN PCB) malfunctioning in the event of a power loss. The circuit may fail to turn off the system main breaker when 2 of the 3 power phases are lost.
-
- Terminated Z-0888-05
Image reconstruction software problem that could potentially cause multiple patient images to be stored in one patient folder.
-
- Terminated Z-0797-05
The electrical cables that supply gradient power to the magnet can become loose over time and create a potential fire hazard.
-
- Terminated Z-0846-05
The absorber unit in the helium compressor which is located in equipment rooms, and is not accessed by patients or system operators, may fail under pressure, causing an explosion that damages the compressor.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 19 Sep 2026
The Hitachi Altaire is a Nuclear Magnetic Resonance Imaging (MRI) System, designed for diagnostic imaging of the human body using strong magnetic fields and radio waves. It generates detailed images of organs, tissues, and skeletal structures to aid in the detection and evaluation of medical conditions, including musculoskeletal, neurological, and cardiovascular disorders.
This MRI system is intended for use in clinical radiology settings and is supplied as a reusable medical device. It operates under FDA 510(k) clearance as a Class 2 device, with regulatory classification under 892.1000. The system is not single-use and requires proper maintenance, calibration, and adherence to MRI safety protocols, including compliance with magnetic field strength and spatial parameters. Storage and handling must follow manufacturer guidelines to ensure optimal performance and patient safety.
Other devices from Hitachi Medical Systems America, Inc.
See all devices from Hitachi Medical Systems America, Inc. →