Hospira Phoenix Infusion System with Hospira Mednet Software
by Abbott Laboratories 510(k)
Active
510(k) K041550
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK041550 | Date30 Jun 2004 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2056-2016
- Terminated Z-2055-2016
Frequent Proximal Occlusion Alarms (POAs) occurring on Symbiq devices with list numbers 16026 and 16027. When there are false alarms, the pump will continue to alarm even though there is no occlusion present. False POAs could result in a delay/interruption in therapy, and require the clinician to reset the alarm multiple times or to replace the pump.
-
- Terminated Z-0069-2013
- Terminated Z-0070-2013
The Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user input could result in a delay in therapy or result in over delivery or under delivery of medication if the user does not confirm the programmed values on the pump's confirmation screen prior to starting the infusion.
-
- Terminated Z-0362-2017
- Terminated Z-0361-2017
May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute audible alarm from the backup buzzer and deactivate the couch screen and the automated loading/unloading of cassettes. Delay or interruption of therapy.
-
- Terminated Z-1895-2010
- Terminated Z-1894-2010
Hospira has received numerous customer complaints of the Symbiq infusion pump's failure to detect air-in-line conditions during operation.
-
- Terminated Z-2381-2010
If the administrator set is removed prior to the cassette carriage fully opening without closing the slide/roller clamp, unrestricted flow and delivery may result.
-
- Terminated Z-2390-2010
- Terminated Z-2391-2010
The clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container.
-
- Terminated Z-0127-2011
- Terminated Z-0128-2011
Hospira has identified motor encoder failures in the Symbiq pumping mechanism that causes the infuser to cease operation during infusion therapy, resulting in delay or interruption of infusion therapy.
-
- Terminated Z-0808-2010
There is a potential for fluid to leak from the backcheck valve which could result in inaccurate delivery and occupational exposure.
-
- Terminated Z-1061-2010
- Terminated Z-1058-2010
- Terminated Z-1059-2010
- Terminated Z-1057-2010
- Terminated Z-1062-2010
- Terminated Z-1051-2010
- Terminated Z-1045-2010
- Terminated Z-1060-2010
- Terminated Z-1053-2010
- Terminated Z-1047-2010
- Terminated Z-1049-2010
- Terminated Z-1046-2010
- Terminated Z-1052-2010
- Terminated Z-1048-2010
- Terminated Z-1050-2010
- Terminated Z-1054-2010
- Terminated Z-1063-2010
- Terminated Z-1056-2010
- Terminated Z-1055-2010
Company has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
-
- Terminated Z-1815-2009
- Terminated Z-1816-2009
Potential delay/underinfusion of critical therapy-- Devices experience increased frequency of alarms that result in temporary pump stoppage.
-
- Terminated Z-0944-2009
Hypersensitivity to normal motor friction/drag may cause malfunction errors and alarms that are designed to stop pump operations.
Specification
- 510(k) Number
- FDA Product Code
- FRN
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent