Human Igg Subclass Liquid Reagent Kits (for Use on Hitachi 911/ 912/ 917 and Modular P)
by The Binding Site, Ltd. 510(k)
Active
510(k) K090920
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK090920 | Date30 Dec 2009 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1070-2017
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
Specification
- 510(k) Number
- FDA Product Code
- CFN
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.5510
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent