Human Igm Liquid Reagent Kit for Use on the Spaplus Analyzer
by The Binding Site, Ltd. 510(k)
Active
510(k) K082129
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK082129 | Date10 Feb 2009 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0604-2019
Firm is offering advice to customers experiencing challenges when validating the calibration curve of the IgM kits. This is indicated by kit control values reporting outside the specified ranges.
-
- Terminated Z-0477-2019
After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been shown to be similarly affected.
Specification
- 510(k) Number
- FDA Product Code
- CFN
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.5510
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent