Genesys HTA System
by Axora Medical, Inc. PMA
Active
PMA P000040
Overview
The Genesys HTA System is a thermal ablation device designed to ablate the endometrial lining of the uterus. It is indicated for use in premenopausal women suffering from menorrhagia, or excessive uterine bleeding, caused by benign conditions where childbearing is complete.
This system is authorized under FDA PMA P000040. It is intended for clinical use by medical professionals to address specific gynecological health needs in patients who have finished childbearing.
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA PMAAPPR | NumberP000040 | Date20 Apr 2001 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA PMA Active
Recalls
-
- Terminated Z-2208-2009
- Terminated Z-2209-2009
Cracked procedure sheaths, incorrect care/use of device, and console malfunctions may result in fluid leaks and loss of a cervical seal that are responsible for burns.
-
- Terminated Z-0688-2007
- Terminated Z-0689-2007
- Terminated Z-0691-2007
- Terminated Z-0687-2007
- Terminated Z-0690-2007
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
-
- Terminated Z-1020-05
The Hysteroscope Adapter used with the HTA System may disassemble and result in a fluid leak.
Specification
- Trade Name
- Hydro Thermablator Endometrial Ablation System
- Generic Name
- Device, thermal ablation, endometrial
- PMA Number
- FDA Product Code
- MNB
- Device Class
- Class III
- Advisory Committee
- Obstetrics/Gynecology
- FDA Decision
- APPR Approved
Other devices from Axora Medical, Inc.
Questions
What is the primary purpose of the Genesys HTA System?
The system is a thermal ablation device intended to ablate the endometrial lining of the uterus to treat menorrhagia, which is excessive uterine bleeding caused by benign conditions.
Who is the intended patient for this device?
The device is intended for premenopausal women who experience menorrhagia due to benign causes and for whom childbearing is complete.
Is the Genesys HTA System FDA authorized?
Yes, the device is authorized under FDA PMA number P000040.