I-Stat Prothrombin Time Test
by I-Stat Corporation 510(k)
Active
510(k) K020355
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK020355 | Date14 May 2002 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2361-2015
Certain lots of PT/INR cartridges may generate a higher than typical number of PT/INR star outs potentially delaying the generation of patient results for PT/INR.
-
- Terminated Z-0451-2014
Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR.
-
- Terminated Z-2157-2012
The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results.
-
- Terminated Z-1996-2010
A high mean bias (12-16%) was documented in 8 lots of i-STAT PT/INR cartridges. This may lead a clinician to believe that a patient is adequately anticoagulated when they are not. Inadequate anticoagluation therapy may be given and issues with clotting and/or thrombosis could occur. No reports of injury have been received.
-
- Terminated Z-0364-2007
Patients treated with the antibiotic CUBICIN can cause a concentration dependent false prolongation of prothrombin time (PT) and elevation of INR (International Normalized Ratio) when using the i-STAT PT/INR cartridges.
Specification
- 510(k) Number
- FDA Product Code
- GJS
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.7750
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent