I.V. Set with Dual-Channel Cassette

by Abbott Laboratories 510(k)

Active United States

510(k) K865060

Regulatory status

as of 26 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK865060 Date6 Feb 1987ClearedStatusActive

Coverage

1 jurisdiction

United States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not known
  • Not covered

United States — FDA 510(k) Active

Recalls

6 recalls · FDA

    • Terminated Z-1491-2012 FDA Not yet classified initiated 25 Apr 2012

    The Lifeshield Primary Plumsets may cause a cassette test failure alarm when the set is loaded onto the infusion pump. A set which results in a cassette test failure alarm may result in delay in therapy, requiring medical intervention.

  • 3 recalls, same reason

    • Terminated Z-0134-06 FDA Unclassified (pre-2012) initiated 29 Sep 2005 via K865060, on the same 510(k)
    • Terminated Z-0133-06 FDA Unclassified (pre-2012) initiated 29 Sep 2005 via K865060, on the same 510(k)
    • Terminated Z-0135-06 FDA Unclassified (pre-2012) initiated 29 Sep 2005 via K865060, on the same 510(k)

    Inlet and outlet tubing on the cassette may be reversed, which would result in reverse flow.

    • Terminated Z-0958-03 FDA Unclassified (pre-2012) initiated 9 Jun 2003

    The drip chambers and/or adapters in the sets were manufactured with non-radiation grade plastic resin, resulting in discoloration of the resin.

    • Terminated Z-0936-03 FDA Unclassified (pre-2012) initiated 13 May 2003 via K865060, on the same 510(k)

    The I.V. administration set incorrectly lists the priming volume from the cassette inlet to distal male adapter as 13.5 mL instead of 11.3 mL

Recalls marked "via" name another device cleared under the same FDA submission as this one.

Specification

510(k)

510(k) Number
FDA Product Code
FRN
Device Class
Class II
Regulation
21 CFR Part 880.5725
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent