# ICU Medical CL2200

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/icu-medical-cl2200>
- Manufacturer: [ICU Medical, Inc.](https://devicerts.com/manufacturers/icu-medical-inc) (United States)
- Device class: Class II (US FDA)
- FDA product code: ONB
- GMDN: Medication transfer set (41222)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 00887709097742 · listed 2021-07-27 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K131549 · decision 2013-07-26 · decision SESE · status: Active · from GUDID

## Recalls

- Z-3159-2026 · Class II · ongoing · initiated 2026-05-12: Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
- Z-3161-2026 · Class II · ongoing · initiated 2026-05-12: Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
- Z-2210-2026 · Class II · ongoing · initiated 2026-03-24: Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy interruption with the potential for the patient to receive less infusion solution than intended. Exposure of hazardous substances to the patient, caregiver, or pharmacy technician is possible.
- Z-0328-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0329-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0330-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0331-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0332-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0333-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0334-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0335-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0336-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0337-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0338-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0339-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0340-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0341-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0342-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0343-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0344-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0345-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0346-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0347-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0348-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0349-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0350-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0351-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0352-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0353-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0354-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0355-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0356-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0357-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0358-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0359-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0360-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0361-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0362-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0363-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0364-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0365-2026 · Class II · ongoing · initiated 2025-09-08: Port weld of drug transfer device may separate or break during use and potentially result in a leak.
- Z-0652-2025 · Class II · ongoing · initiated 2024-11-06: Due to finished goods incorrectly packaged and labeled.
- Z-1681-2022 · Class II · ongoing · initiated 2022-06-27: Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.
- Z-1570-2021 · Class II · ongoing · initiated 2021-03-16: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Z-1571-2021 · Class II · ongoing · initiated 2021-03-16: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Z-1572-2021 · Class II · ongoing · initiated 2021-03-16: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Z-1573-2021 · Class II · ongoing · initiated 2021-03-16: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Z-1574-2021 · Class II · ongoing · initiated 2021-03-16: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Z-1575-2021 · Class II · ongoing · initiated 2021-03-16: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
- Z-0703-2021 · Class II · ongoing · initiated 2020-08-31: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
- Z-1096-2019 · Class I · terminated · initiated 2019-02-21 · matched through shared submission K131549, may concern other models: There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
- Z-0805-2019 · Class I · terminated · initiated 2019-01-03: There is a potential for one lot of ChemoLock Vial Spike to contain burr particulate
- Z-1254-2016 · Class II · terminated · initiated 2016-02-25: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.
- Z-1255-2016 · Class II · terminated · initiated 2016-02-25: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.
- Z-1256-2016 · Class II · terminated · initiated 2016-02-25: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.
- Z-1257-2016 · Class II · terminated · initiated 2016-02-25: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.
- Z-1258-2016 · Class II · terminated · initiated 2016-02-25: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.
- Z-1259-2016 · Class II · terminated · initiated 2016-02-25: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.
- Z-1260-2016 · Class II · terminated · initiated 2016-02-25: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.
- Z-1261-2016 · Class II · terminated · initiated 2016-02-25: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.

## Identity

- Trade name: ICU Medical [FDA GUDID (AccessGUDID)]
- Generic name: Medication transfer set [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: CL2200 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 00887709097742 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K131549 [FDA GUDID (AccessGUDID)]
- FDA product code: ONB [FDA GUDID (AccessGUDID)]
- GMDN term: Medication transfer set (41222) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 880.5440 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Supplied sterile: Yes [FDA GUDID (AccessGUDID)]
- Single use: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Expiration date: Yes [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: ICU Medical CL2200, ICU Medical, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/icu-medical-cl2200 (data as of 2026-11-23).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/icu-medical-cl2200.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.