  "format": "devicerts.device.v1",
  "cite_as": "iLet ACE Pump, Beta Bionics, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/ilet-ace-pump-qfg (data as of 2026-11-21).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/ilet-ace-pump-qfg",
    "json_url": "https://devicerts.com/devices/ilet-ace-pump-qfg.json",
    "markdown_url": "https://devicerts.com/devices/ilet-ace-pump-qfg.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-21",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/ilet-ace-pump-qfg",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "iLet ACE Pump",
    "registry_name": "iLet ACE Pump",
    "url": "https://devicerts.com/devices/ilet-ace-pump-qfg",
    "manufacturer": {
      "name": "Beta Bionics, Inc.",
      "url": "https://devicerts.com/manufacturers/beta-bionics-inc",
      "country": "United States",
      "country_basis": "FDA 510(k) applicant",
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K252770",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "QFG"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K252770",
      "number_label": null,
      "device_udi_di": null,
      "date": "2025-09-29",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1996-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-03-12",
      "reason": "insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous\nGlucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for\nsensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.",
      "applies": "this device",
      "via": "K252770"
    },
    {
      "number": "Z-2513-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-09-23",
      "reason": "The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen  transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.",
      "applies": "this device",
      "via": "K252770"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K252770",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "QFG",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 880.5730",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Clinical Chemistry",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}