- **510(k)** · FDA 510(k) · United States · K162314 · decision 2018-04-24 · decision SESE · status: Active

## Recalls

- Z-1590-2026 · Class II · ongoing · initiated 2026-02-03: There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
- Z-1264-2021 · Class II · terminated · initiated 2021-02-11: Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory
- Z-2972-2020 · Class II · ongoing · initiated 2020-08-24: The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.
- Z-0838-2020 · Class II · ongoing · initiated 2019-08-01: The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.

## Identifiers

- 510(k) Number: K162314 [FDA 510(k)]
- FDA product code: JAK [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.1750 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: IMACTIS CT-Navigation system, Imactis SAS. deviCERTS medical device catalogue. https://devicerts.com/devices/imactis-ct-navigation-system-jak (data as of 2026-11-26).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-26
- JSON: <https://devicerts.com/devices/imactis-ct-navigation-system-jak.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/imactis-ct-navigation-system-jak
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.