Immulite 2000 Automated Immunoassay Analyzer
by Dpc Cirrus 510(k)
Active
510(k) K970227
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK970227 | Date8 Apr 1997 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2470-2020
- Terminated Z-2469-2020
Flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard
-
- Terminated Z-1663-2015
- Terminated Z-1661-2015
- Terminated Z-1660-2015
- Terminated Z-1658-2015
- Terminated Z-1662-2015
The bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.
-
- Terminated Z-1832-2014
The keyboard and monitor may not be properly secured to the monitor arm.
-
- Terminated Z-1316-2014
There is a potential for incorrect results to be reported from processing of a sample tube.
-
- Terminated Z-1033-2010
The recall was initiated because Siemens Healthcare Diagnostics has identified there is the potential for patient results to be biased low (-0.3 Sample to Cutoff [S/CO] ratio units) due to incompatibility of the Toxoplasma IgM kits with this particular lot of IgG/IgM Sample Diluent. All available lots of the IMMULITE 2000/2500 Toxoplasma IgM kits (L2KTM/L5KTM) are incompatible with Diluent lot 150
-
- Terminated Z-1891-2010
These substrate lots have shown a tendency to introduce a shift in both control and patient results with multiple methods.
-
- Terminated Z-1399-2009
Design Flaw: The redesign of the Digital Fluidics PCB used in IMMULITE 2000 and IMMULITE 2500 is missing a jumper.
-
- Terminated Z-0994-05
The Transport Chain is used to move a Reaction Tube within the Immulite 2000. It goes on to state the error condition that occurs due to a false trip of the home sensor.
Specification
- 510(k) Number
- FDA Product Code
- JJE
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.2160
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent