Immulite 2000 Specific Allergens
510(k) K021206
Active
510(k)
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK021206 | Date30 May 2002 | StatusActive |
Recalls
-
- Ongoing Z-0491-2026
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DHB
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.5750
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent