Immulite Cea
510(k) K970877
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK970877 | Date22 Sep 1997 | StatusActive |
Recalls
-
- Terminated Z-1678-2018
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
-
- Terminated Z-1015-05
Adverse trend in stability, controls biased or out of range.
-
- Terminated Z-0972-04
Instabililty
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DHX
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.6010
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent