Immulite Intact Pth, Immulite 2000 Intact Pth, Models LKPP1 (100 Tests), LKPP5 (500 Tests), L2KPP2 (200 Tests), L2KPP6
510(k) K013566
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK013566 | Date20 Nov 2001 | StatusActive |
Recalls
-
- Ongoing Z-0366-2026
The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.
-
- Terminated Z-0979-2017
Siemens is conducting this recall for the IMMULITE 2000/IMMULITE 2000 XPi Intact PTH (Intact Parathyroid Hormone) (iPTH) assay kit lot 320. Siemens has confirmed that IMMULITE 2000/IMMULITE 2000 XPi Intact PTH kit lot 320 can exhibit an average negative bias of up to -39% at iPTH concentrations <20 pg/mL with serum and EDTA patient samples vs. a reference kit lot.
-
- Terminated Z-0724-04
High bias noted in samples collected with EDTA. Indication is to not run samples with EDTA.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CEW
- Models / Variants
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1545
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent