Imp. Epikut MT16° Ø4, 5X10, 0mm Ilm 4510
by S.I.N. - Sistema DE Implante Nacional S.A.
Ref ILM 4510Model DENTAL IMPLANTS WITHOUT HYDROXYAPATITE
Active
CE MDR
Valid until
Not listed
FDA recalls
1 ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number999520000001JG 07899995280220 | Date18 Aug 2026 | StatusActive |
| United States | GUDID listing GUDID | Number07899995280220 | Date14 Jun 2023 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-0920-2025
Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade Name
- Imp. Epikut S MT16° Ø4.5x10mm EUDAMEDImp. Epikut MT16° Ø4,5x10,0mm GUDID
- Generic Name
- Screw endosteal dental implant, one-piece
- Device Name
- Implant Tryon | Implant Strong Sw | Implant Epikut
- Model / Ref / Catalog
- ILM 4510, DENTAL IMPLANTS WITHOUT HYDROXYAPATITE
- Primary DI / UDI-DI
- 07899995280220
- Basic UDI-DI
- 999520000001JG
- 510(k) Number
- FDA Product Code
- DZE
- EMDN Code
- P01020101 Dental implants
- GMDN Term
- Screw endosteal dental implant, one-piece
- Device Class
- Class IIb EUDAMEDClass II GUDID
- Regulation Number
- 872.3640
- EU market of registration
- Portugal
- GUDID distribution status
- In Commercial Distribution
- MRI Safety
- MR Conditional
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class IIb | Class II |
Other devices from S.I.N. - Sistema DE Implante Nacional S.A.
See all devices from S.I.N. - Sistema DE Implante Nacional S.A. →
Properties
Other properties (16)
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No