{
  "format": "devicerts.device.v1",
  "cite_as": "Impella 0043-0003, Abiomed, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/impella-0043-0003 (data as of 2026-11-27).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/impella-0043-0003",
    "json_url": "https://devicerts.com/devices/impella-0043-0003.json",
    "markdown_url": "https://devicerts.com/devices/impella-0043-0003.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-27",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/impella-0043-0003",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Impella 0043-0003",
    "registry_name": "Impella 0043-0003",
    "url": "https://devicerts.com/devices/impella-0043-0003",
    "manufacturer": {
      "name": "Abiomed, Inc.",
      "url": "https://devicerts.com/manufacturers/abiomed-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class III",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Catalogue number",
        "values": [
          {
            "value": "0043-0003",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "00813502011135",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "PMA number",
        "values": [
          {
            "value": "P140003",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "P170011",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "OZD",
      "PYX"
    ],
    "emdn": [],
    "gmdn": [
      {
        "code": "57791",
        "term": "Infusion pump cassette"
      }
    ],
    "description": null,
    "noindex": true
  },
  "authorizations": [
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA PMA",
      "number": "P170011",
      "number_label": null,
      "device_udi_di": null,
      "date": "2017-09-20",
      "date_kind": "decision",
      "decision_code": "APPR",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA PMA",
      "number": "P140003",
      "number_label": null,
      "device_udi_di": null,
      "date": "2015-03-23",
      "date_kind": "decision",
      "decision_code": "APPR",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "00813502011135",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2014-07-31",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1589-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-02-18",
      "reason": "Increased risk of purge leaks with Generation 1 purge cassettes.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0208-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-09-18",
      "reason": "Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s):\r\n1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. \r\n2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. \r\n3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. \r\n4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0209-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-09-18",
      "reason": "Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s):\r\n1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. \r\n2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. \r\n3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. \r\n4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0210-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-09-18",
      "reason": "Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s):\r\n1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. \r\n2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. \r\n3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. \r\n4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-3148-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-08-12",
      "reason": "The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-3149-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-08-12",
      "reason": "The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-3150-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-08-12",
      "reason": "The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-2457-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-05-18",
      "reason": "Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-2360-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-05-14",
      "reason": "Potential software error in the Automated Impella Controller (\"AIC\" ) when used in conjunction with left ventricular Impella devices.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-2211-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-04-20",
      "reason": "A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates.\nIssue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls.\nIssue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues.\nIssue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure.\nIssue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-1591-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-02-16",
      "reason": "Potential that the \"Purge System Blocked\"  alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-1469-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-01-27",
      "reason": "Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.",
      "applies": "this device",
      "via": "P170011"
    },
    {
      "number": "Z-1470-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-01-27",
      "reason": "Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.",
      "applies": "this device",
      "via": "P170011"
    },
    {
      "number": "Z-1471-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-01-27",
      "reason": "Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.",
      "applies": "this device",
      "via": "P170011"
    },
    {
      "number": "Z-1030-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-12-04",
      "reason": "Device packaged in incorrect outer box carton.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0069-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-10-01",
      "reason": "Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0136-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-09-16",
      "reason": "Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-2164-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-06-23",
      "reason": "A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0829-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-12-12",
      "reason": "Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.",
      "applies": "this device",
      "via": "P170011"
    },
    {
      "number": "Z-0830-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-12-12",
      "reason": "Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.",
      "applies": "this device",
      "via": "P170011"
    },
    {
      "number": "Z-2319-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-05-31",
      "reason": "Nine (9) Impella CP pumps failed inspection and were inadvertently released.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0980-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-12-27",
      "reason": "IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0636-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-06",
      "reason": "The following features have been disabled from the web-based portal because the FDA has not evaluated these features for safety and effectiveness:\n\nNotifications via email feature; AIC alarm color on case tile feature; Sort case tiles by AIC alarm color feature; Pump metric display on case tile feature",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0440-2024",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2023-10-10",
      "reason": "Pump not detected as connected to controller due to software issue.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0441-2024",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2023-10-10",
      "reason": "Pump not detected as connected to controller due to software issue.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-1590-2023",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-04-17",
      "reason": "The Impella 5.5 pump experienced a heightened complaint rate for purge leaks.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-1175-2024",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2021-03-01",
      "reason": "Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-0345-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-09-04",
      "reason": "This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-3251-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2019-02-19",
      "reason": "Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure\nsignal the AIC system performance depends on. \n2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key\nduring disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.",
      "applies": "this device",
      "via": "P140003"
    },
    {
      "number": "Z-2238-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2011-01-17",
      "reason": "Retrospective submission for following issues identified: \n1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. \n2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.",
      "applies": "this device",
      "via": "P140003"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "Impella",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Infusion pump cassette",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / Ref / Catalog",
          "values": [
            {
              "value": "0043-0003",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "00813502011135",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "PMA number",
          "values": [
            {
              "value": "P140003",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "P170011",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "OZD",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "PYX",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Infusion pump cassette",
              "code": "57791",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class III",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "870.4360",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}
