Skip to content

Impella 5.0 AIC Pump Set

by Abiomed, Inc.

Impella 5.0 AIC Pump Set (005062), a medical device (FDA Class II) made by Abiomed, Inc., was listed in FDA GUDID on 31 Jul 2014 (UDI-DI 00813502011180); the registry lists it as not in commercial distribution (GUDID) as of 22 Oct 2026. FDA recalls on record: 3.

Listed FDA GUDID listing, not in commercial distribution

Valid until

Not applicable

Database listing, no authorization

FDA recalls

3 ongoing

This device is named in the recall

Documents

None

No authorization documents

Trust

Unverified

Source GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00813502011180 UDI-DI Date31 Jul 2014ListedStatusListed GUDID: not in commercial distribution

Official records

From the FDA record

Device description
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Recalls

3 recalls · 3 ongoing · FDA

    • Ongoing Z-0980-2024 FDA Class I initiated 27 Dec 2023 UDI-DI 00813502011180 listed

    IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.

    • Ongoing Z-1074-2024 FDA Class II initiated 27 Dec 2023 UDI-DI 00813502011180 listed

    New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.

    • Ongoing Z-2112-2023 FDA Class I initiated 14 Jun 2023 UDI-DI 00813502011180 listed

    There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.

Timeline

4 dated events

  1. Listed FDA GUDID 00813502011180
  2. Recall initiated FDA recalls Z-2112-2023
  3. Recall initiated FDA recalls Z-0980-2024
  4. Recall initiated FDA recalls Z-1074-2024

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — GUDID listing Listed

Specification

GUDID

Trade name
Impella
Generic name
Intracardiac circulatory assist axial-pump catheter
Model / Ref / Catalog
Impella 5.0 AIC Pump Set, 005062
Primary DI / UDI-DI
00813502011180
FDA product code
KFM, OZD
GMDN term
Intracardiac circulatory assist axial-pump catheter
Device class
Class II
Regulation number
870.4360
GUDID distribution status
Not in Commercial Distribution

Similar devices

Other devices filed under FDA product code KFM (Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type).

Other devices from Abiomed, Inc.

See all devices from Abiomed, Inc. →

Properties

  • Supplied sterile
  • Single use
  • Prescription use
  • Convenience kit
  • Lot / batch number
  • Serial number
  • Expiration date
Other properties (8, all answered No)