# Impella2.5, ImpellaCP, Impella CP with SmartAssist

> Active: European Union and United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/impella-2-5-system-p140003>
- Manufacturer: [Abiomed, Inc.](https://devicerts.com/manufacturers/abiomed-inc) (United States)
- Device class: Class III (US FDA)
- FDA product code: OZD

## Overview

*AI-generated summary, 2026-09-18. Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data is the record.*

The Impella 2.5 and Impella CP are temporary non-roller type left heart support blood pumps designed to provide ventricular support. These systems are indicated for use during high-risk percutaneous coronary interventions in hemodynamically stable patients with severe coronary artery disease and depressed left ventricular ejection fraction.

These devices are intended for use when a heart team has determined that high-risk percutaneous coronary intervention is the appropriate therapeutic option. The systems are designed to prevent hemodynamic instability during planned temporary coronary occlusions and may reduce peri- and post-procedural adverse events. The Impella 2.5 system is indicated for temporary use for up to 6 hours.

## Registrations and authorizations

- **PMA** · FDA PMA · United States · P140003 · decision 2015-03-23 · decision APPR · status: Active
- **EUDAMED** · CE/MDR · European Union · status: Active
- **EUDAMED** · CE/MDD · European Union · status: Active

## Recalls

- Z-0208-2027 ·  · ongoing · initiated 2026-09-18: Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
- Z-0209-2027 ·  · ongoing · initiated 2026-09-18: Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
- Z-0210-2027 ·  · ongoing · initiated 2026-09-18: Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
- Z-3148-2026 · Class I · ongoing · initiated 2026-08-12: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
- Z-3149-2026 · Class I · ongoing · initiated 2026-08-12: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
- Z-3150-2026 · Class I · ongoing · initiated 2026-08-12: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
- Z-2457-2026 · Class I · ongoing · initiated 2026-05-18: Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
- Z-2360-2026 · Class I · ongoing · initiated 2026-05-14: Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
- Z-2211-2026 · Class I · ongoing · initiated 2026-04-20: A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.
- Z-1589-2026 · Class I · ongoing · initiated 2026-02-18: Increased risk of purge leaks with Generation 1 purge cassettes.
- Z-1591-2026 · Class I · ongoing · initiated 2026-02-16: Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008\*) pumps.
- Z-1030-2026 · Class II · ongoing · initiated 2025-12-04: Device packaged in incorrect outer box carton.
- Z-0069-2026 · Class I · ongoing · initiated 2025-10-01: Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
- Z-0136-2026 · Class I · ongoing · initiated 2025-09-16: Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.
- Z-2164-2025 · Class I · ongoing · initiated 2025-06-23: A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected
- Z-2319-2024 · Class I · ongoing · initiated 2024-05-31: Nine (9) Impella CP pumps failed inspection and were inadvertently released.
- Z-0980-2024 · Class I · ongoing · initiated 2023-12-27: IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.
- Z-0636-2024 · Class II · ongoing · initiated 2023-12-06 · matched through shared submission P140003, may concern other models: The following features have been disabled from the web-based portal because the FDA has not evaluated these features for safety and effectiveness: Notifications via email feature; AIC alarm color on case tile feature; Sort case tiles by AIC alarm color feature; Pump metric display on case tile feature
- Z-0440-2024 · Class II · terminated · initiated 2023-10-10: Pump not detected as connected to controller due to software issue.
- Z-0441-2024 · Class II · terminated · initiated 2023-10-10: Pump not detected as connected to controller due to software issue.
- Z-1590-2023 · Class I · ongoing · initiated 2023-04-17: The Impella 5.5 pump experienced a heightened complaint rate for purge leaks.
- Z-1175-2024 · Class II · terminated · initiated 2021-03-01: Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
- Z-0345-2020 · Class II · terminated · initiated 2019-09-04: This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.
- Z-3251-2026 · Class II · ongoing · initiated 2019-02-19: Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
- Z-2238-2026 · Class I · ongoing · initiated 2011-01-17: Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.

## Identity

- Trade name: IMPELLA 2.5 System [FDA PMA]
- Generic name: Temporary non-roller type left heart support blood pump [FDA PMA]

## Identifiers

- PMA number: P140003 [FDA PMA]
- FDA product code: OZD [FDA PMA]

## Classification

- Device class: Class III [FDA PMA]
- Advisory committee: Cardiovascular [FDA PMA]
- FDA decision: APPR [FDA PMA]

## Source and citation

- Cite as: Impella2.5, ImpellaCP, Impella CP with SmartAssist, Abiomed, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/impella-2-5-system-p140003 (data as of 2026-11-27).
- Data sources: [FDA PMA](https://open.fda.gov/apis/device/pma/), [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-06
- Statuses computed as of: 2026-11-27
- JSON: <https://devicerts.com/devices/impella-2-5-system-p140003.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/impella-2-5-system-p140003
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.
