Imri 1.5T and Imri 3T S
by Imris, Inc. 510(k)
Active Open recall
510(k) K133692
Regulatory status
Recall ongoing. 4 ongoing FDA recalls name this device. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK133692 | Date10 Feb 2014 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-0164-2026
- Ongoing Z-0163-2026
- Ongoing Z-0161-2026
- Ongoing Z-0162-2026
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
- Terminated Z-1645-2018
The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.
-
- Terminated Z-1098-2016
Emergency helium venting lines (i.e. quench lines) may not have been properly installed.
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent