# iMRX System iMRX System AR-DR MR30RS

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/imrx-system-imrx-system-ar-dr-mr>
- Manufacturer: [IMRIS Imaging, Inc.](https://devicerts.com/manufacturers/imris-imaging-inc) (United States)
- Device class: Class II (US FDA)
- FDA product code: IZI, LNH
- GMDN: Full-body MRI system, superconducting magnet (37654)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 00857534006189 · listed 2016-09-23 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K101813 · decision 2010-10-05 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0161-2026 · Class I · ongoing · initiated 2025-09-30: There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Z-0162-2026 · Class I · ongoing · initiated 2025-09-30: There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Z-0163-2026 · Class I · ongoing · initiated 2025-09-30: There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Z-0164-2026 · Class I · ongoing · initiated 2025-09-30: There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Z-0762-2020 · Class II · terminated · initiated 2019-11-26: There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible

## Identity

- Trade name: iMRX System [FDA GUDID (AccessGUDID)]
- Generic name: Full-body MRI system, superconducting magnet [FDA GUDID (AccessGUDID)]
- Model / reference: iMRX System AR-DR MR30RS [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 00857534006189 [FDA GUDID (AccessGUDID)]
- FDA product code: IZI [FDA GUDID (AccessGUDID)]; LNH [FDA GUDID (AccessGUDID)]
- GMDN term: Full-body MRI system, superconducting magnet (37654) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 892.1600 [FDA GUDID (AccessGUDID)]; 892.1000 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Clinical & Handling

- MRI safety: MR Conditional [FDA GUDID (AccessGUDID)]

## Properties

- Contains latex: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Manufacturing date: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: iMRX System iMRX System AR-DR MR30RS, IMRIS Imaging, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/imrx-system-imrx-system-ar-dr-mr (data as of 2026-11-23).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/imrx-system-imrx-system-ar-dr-mr.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/imrx-system-imrx-system-ar-dr-mr
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.