Infinity™ 6660
Infinity 6660 is a deep brain electrical stimulation system (FDA Class III; FDA product code MHY, Stimulator, Electrical, Implanted, for Parkinsonian Tremor) made by St. Jude Medical, Inc., approved under FDA PMA P140009 on 12 Jun 2015 and listed in FDA GUDID since 30 Sep 2016 (UDI-DI 05415067020246), active as of 23 Oct 2026. 7 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number05415067020246 | Date30 Sep 2016 | StatusListed |
| United States | FDA PMAAPPR | NumberP140009 | Date12 Jun 2015 | StatusActive |
Official records
- AccessGUDID record 05415067020246
- FDA PMA database P140009
From the FDA record
- Device description
- Implantable Pulse Generator
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA PMA P140009 decided 12 Jun 2015 Active
- FDA GUDID UDI-DI 05415067020246 listed 30 Sep 2016 Listed
Recalls
-
- Ongoing Z-0459-2026
There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).
-
- Ongoing Z-2429-2024
Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.
-
- Terminated Z-0023-2018
Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.
-
- Terminated Z-2693-2017
- Terminated Z-2694-2017
- Terminated Z-2695-2017
- Terminated Z-2696-2017
The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
Timeline
- Approved FDA PMA P140009
- Listed FDA GUDID 05415067020246
- Recall initiated FDA recalls Z-2693-2017
- Recall initiated FDA recalls Z-2694-2017
- Recall initiated FDA recalls Z-2695-2017
- Recall initiated FDA recalls Z-2696-2017
- Recall initiated FDA recalls Z-0023-2018
- Recall initiated FDA recalls Z-2429-2024
- Recall initiated FDA recalls Z-0459-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade name
- Infinity™
- Generic name
- Deep brain electrical stimulation system
- Model / Ref / Catalog
- 6660
- Primary DI / UDI-DI
- 05415067020246
- PMA number
- FDA product code
- MHY
- GMDN term
- Deep brain electrical stimulation system
- Device class
- Class III
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code MHY (STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR).