Infinix INFX-8000V, V6.40

by Toshiba Medical Systems Coroporation

510(k) K170909

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

4 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK170909 Date16 Jun 2017ClearedStatusActive

Recalls

5 recalls · 4 ongoing · FDA

    • Ongoing Z-1666-2022 FDA Class II initiated 31 Jan 2022

    The fluoroscopic dose rate might exceed the conforming value during biplane fluoroscopy.

    • Ongoing Z-1174-2021 FDA Class II initiated 5 Feb 2021

    The x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.

    • Ongoing Z-1269-2021 FDA Class II initiated 8 Dec 2020

    System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.

    • Ongoing Z-2325-2020 FDA Class II initiated 3 Mar 2020

    Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.

    • Terminated Z-0844-2018 FDA Class II initiated 29 Dec 2017

    During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
OWB
Device Class
Class II
Regulation
21 CFR Part 892.1650
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Toshiba Medical Systems Coroporation