Inflamma Dry
by Rapid Pathogen Screening, Inc.
510(k) K132066
Active
510(k)
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK132066 | Date21 Nov 2013 | StatusActive |
Recalls
-
- Ongoing Z-1636-2024
- Ongoing Z-1637-2024
There are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned. In instances of positive test results, the positivite test line may not be visible/ hidden under the cassette of the test device thus potentially leading to a false negative result.
-
- Terminated Z-2803-2017
In-house testing revealed the presence of plastic spurs at the tips of some Sample Collectors.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- PFQ
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1540
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent