Infx-8000F (Infinix Csi and Infinix Dpi)
by Toshiba America Medical Systems, In.C
510(k) K081624
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK081624 | Date23 Jun 2008 | StatusActive |
Recalls
-
- Ongoing Z-1176-2021
The x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.
-
- Ongoing Z-1271-2021
System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
-
- Ongoing Z-2327-2020
Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
-
- Terminated Z-0360-2020
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.
-
- Terminated Z-0842-2018
During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.
-
- Terminated Z-0994-2013
Lost and incorrect images. When using the "F-Rec" fluoroscopy image acquisition recording option in the "manual mode" an infrequent software error may cause images from another study to be displayed on the monitor. The acquired images may be lost requiring additional image acquisitions to be performed.
-
- Terminated Z-0331-2012
The recall was initiated because Toshiba has confirmed that during patient transfer from a gurney to the catherization table, the tabletop may rotate unexpectedly when sufficient force is applied to the tabletop. The result of the tabletop rotating unexpectedly is the patient being tranferred may fall from the tabletop.
-
- Terminated Z-0005-2009
Audible signal is not available at all times when using the HLC fluoroscopy mode. Special means of activation are not designed in for selection of High Level Controls, as required by regulation.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MQB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1680
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent