Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei

by Philips Medical Systems Nederland B.V.

510(k) K213516

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

45 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Abbreviated NumberK213516 Date3 Mar 2022ClearedStatusActive

Recalls

45 recalls · 45 ongoing · FDA

  • 3 recalls, same reason

    • Ongoing Z-3311-2026 FDA Class II initiated 28 Aug 2026
    • Ongoing Z-3316-2026 FDA Class II initiated 28 Aug 2026
    • Ongoing Z-3313-2026 FDA Class II initiated 28 Aug 2026

    When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

  • 3 recalls, same reason

    • Ongoing Z-3243-2026 FDA Class II initiated 21 Aug 2026
    • Ongoing Z-3239-2026 FDA Class II initiated 21 Aug 2026
    • Ongoing Z-3242-2026 FDA Class II initiated 21 Aug 2026

    A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

  • 2 recalls, same reason

    • Ongoing Z-1969-2026 FDA Class II initiated 14 Apr 2026
    • Ongoing Z-1963-2026 FDA Class II initiated 14 Apr 2026

    The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

  • 2 recalls, same reason

    • Ongoing Z-1213-2026 FDA Class II initiated 3 Dec 2025
    • Ongoing Z-1214-2026 FDA Class II initiated 3 Dec 2025

    The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

  • 2 recalls, same reason

    • Ongoing Z-2207-2025 FDA Class II initiated 30 Jun 2025
    • Ongoing Z-2214-2025 FDA Class II initiated 30 Jun 2025

    The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

  • 2 recalls, same reason

    • Ongoing Z-1805-2025 FDA Class II initiated 13 May 2025
    • Ongoing Z-1792-2025 FDA Class II initiated 13 May 2025

    Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.

  • 6 recalls, same reason

    • Ongoing Z-1782-2025 FDA Class II initiated 24 Mar 2025
    • Ongoing Z-1776-2025 FDA Class II initiated 24 Mar 2025
    • Ongoing Z-1783-2025 FDA Class II initiated 24 Mar 2025
    • Ongoing Z-1772-2025 FDA Class II initiated 24 Mar 2025
    • Ongoing Z-1770-2025 FDA Class II initiated 24 Mar 2025
    • Ongoing Z-0951-2024 FDA Class II initiated 3 Jan 2024 on shared 510(k) K213516

    The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

  • 13 recalls, same reason

    • Ongoing Z-3201-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3193-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3198-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3199-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3205-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3202-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3196-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3200-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3203-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3194-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3204-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3195-2024 FDA Class II initiated 3 Sep 2024
    • Ongoing Z-3197-2024 FDA Class II initiated 3 Sep 2024

    Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

  • 3 recalls, same reason

    • Ongoing Z-1570-2024 FDA Class II initiated 12 Mar 2024
    • Ongoing Z-1571-2024 FDA Class II initiated 12 Mar 2024
    • Ongoing Z-1588-2024 FDA Class II initiated 12 Mar 2024

    Patient support table floor plate may be incorrectly installed.

  • 2 recalls, same reason

    • Ongoing Z-2395-2023 FDA Class II initiated 12 Jun 2023
    • Ongoing Z-2396-2023 FDA Class II initiated 12 Jun 2023

    The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.

  • 2 recalls, same reason

    • Ongoing Z-2399-2023 FDA Class II initiated 12 Jun 2023
    • Ongoing Z-2398-2023 FDA Class II initiated 12 Jun 2023

    During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.

  • 5 recalls, same reason

    • Ongoing Z-0463-2023 FDA Class II initiated 22 Nov 2022
    • Ongoing Z-0461-2023 FDA Class II initiated 22 Nov 2022
    • Ongoing Z-0462-2023 FDA Class II initiated 22 Nov 2022
    • Ongoing Z-0459-2023 FDA Class II initiated 22 Nov 2022
    • Ongoing Z-0460-2023 FDA Class II initiated 22 Nov 2022

    Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LNH
Device Class
Class II
Regulation
21 CFR Part 892.1000
Advisory Committee
Radiology
FDA Decision