Ingenia Elition S and Ingenia Elition X R5.4
by Philips Medical Systems Nederland B.V.
510(k) K173356
Active
Valid until
No expiry
FDA recalls
10 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK173356 | Date6 Mar 2018 | StatusActive |
Recalls
-
- Ongoing Z-2395-2023
- Ongoing Z-2396-2023
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
-
- Ongoing Z-2398-2023
- Ongoing Z-2399-2023
During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.
-
- Ongoing Z-0460-2023
- Ongoing Z-0461-2023
- Ongoing Z-0463-2023
- Ongoing Z-0462-2023
- Ongoing Z-0459-2023
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
-
- Ongoing Z-0974-2024
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
-
- Terminated Z-0922-2019
- Terminated Z-0923-2019
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
-
- Terminated Z-0694-2019
During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent