Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems
by Philips Medical Systems Nederland B.V.
510(k) K183063
Active
Valid until
No expiry
FDA recalls
25 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK183063 | Date14 Feb 2019 | StatusActive |
Recalls
-
- Ongoing Z-3309-2026
- Ongoing Z-3310-2026
- Ongoing Z-3308-2026
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
-
- Ongoing Z-3239-2026
- Ongoing Z-3238-2026
- Ongoing Z-3237-2026
- Ongoing Z-3236-2026
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
-
- Ongoing Z-1956-2026
- Ongoing Z-1957-2026
- Ongoing Z-1965-2026
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
-
- Ongoing Z-1216-2026
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
-
- Ongoing Z-1788-2025
- Ongoing Z-1789-2025
- Ongoing Z-1791-2025
- Ongoing Z-1790-2025
Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
-
- Ongoing Z-2396-2023
- Ongoing Z-2395-2023
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
-
- Ongoing Z-2398-2023
- Ongoing Z-2399-2023
During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.
-
- Ongoing Z-0459-2023
- Ongoing Z-0460-2023
- Ongoing Z-0461-2023
- Ongoing Z-0463-2023
- Ongoing Z-0462-2023
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
-
- Ongoing Z-0975-2024
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent