Innova Igs 5 001
Listed
Valid until
Not applicable
FDA recalls
6 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00840682124621 | Date24 Jan 2018 | StatusListed |
Recalls
-
- Ongoing Z-1783-2024
- Ongoing Z-1778-2024
- Ongoing Z-1782-2024
- Ongoing Z-1777-2024
GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.
-
- Ongoing Z-1200-2019
- Ongoing Z-1201-2019
There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
-
- Terminated Z-1864-2021
The IGS system can experience a single vertical line defect where the vertical line divides and horizontally shifts live monitor images into two unequal image parts within the monitor display screen.
-
- Terminated Z-2386-2018
- Terminated Z-2387-2018
The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use.
-
- Terminated Z-0763-2016
GE Healthcare has recently become aware of a reported incident in which a Mavig suspension arm with monitor assembly fell to the floor.
-
- Terminated Z-0413-2016
GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of displayed imaging (loss of monitor video) involving Optima CL323i & Optima IGS 320 systems .
-
- Terminated Z-0414-2016
GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of displayed imaging (loss of monitor video) involving Optima CL323i & Optima IGS 320 systems.
-
- Terminated Z-1976-2015
- Terminated Z-1974-2015
- Terminated Z-1973-2015
- Terminated Z-1979-2015
- Terminated Z-1981-2015
- Terminated Z-1977-2015
- Terminated Z-1985-2015
- Terminated Z-1983-2015
- Terminated Z-1980-2015
- Terminated Z-1982-2015
- Terminated Z-1984-2015
- Terminated Z-1978-2015
GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery).
-
- Terminated Z-0451-2015
Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
-
- Terminated Z-0853-2014
GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the user changes the size of the lateral FOV (Field of View) and releases the biplane footswitch pedal simultaneously. This could lead to a system lock up requiring
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Innova
- Generic Name
- Stationary angiographic x-ray system, digital
- Model / Reference
- IGS 5 001
- Primary DI / UDI-DI
- 00840682124621
- FDA Product Code
- JAA, OWB, IZI
- GMDN Term
- Stationary angiographic x-ray system, digital
- Device Class
- Class II
- Regulation Number
- 892.1650, 892.1600
- GUDID distribution status
- Not in Commercial Distribution
Other devices from GE Medical Systems, LLC
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No