INRatio 2 PT Monitoring System

by Hemosense, Inc.

510(k) K072727

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

6 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK072727 Date26 Oct 2007ClearedStatusActive

Recalls

6 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-2357-2016 FDA Class I initiated 11 Jul 2016
    • Terminated Z-2360-2016 FDA Class I initiated 11 Jul 2016

    Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

  • 2 recalls, same reason

    • Terminated Z-0880-2015 FDA Class I initiated 5 Dec 2014
    • Terminated Z-0882-2015 FDA Class I initiated 5 Dec 2014

    In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

    • Terminated Z-0188-2012 FDA Unclassified (pre-2012) initiated 11 Oct 2011 on shared 510(k) K072727

    Alere has initiated a recall of the Professional Disposables International (PDI) alcohol prep products due to the potential presence of low levels of the microorganism Bacillus cereus.

    • Terminated Z-1256-2009 FDA Unclassified (pre-2012) initiated 23 Dec 2008

    INRatio Monitors that were manufactured with a specific EEPROM memory chip fail during use. The monitors report an error code and do not report a test result.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
GJS
Device Class
Class II
Regulation
21 CFR Part 864.7750
Advisory Committee
Hematology
FDA Decision
SESE Substantially Equivalent

Overview

AI-generated summary, 20 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The INRatio 2 PT Monitoring System is a point-of-care device designed for the measurement of prothrombin time (PT) and derived international normalized ratio (INR) values. It is classified as a blood coagulation test system, enabling rapid and accurate monitoring of patients on anticoagulant therapy, particularly those using vitamin K antagonists. The system provides real-time results, facilitating immediate clinical decision-making in settings such as outpatient clinics, anticoagulation management services, and home monitoring programs.

The device operates using a portable, handheld analyzer that requires a small blood sample applied to a test cartridge. Supplied as a single-use system, it does not require external calibration or additional control tests, as internal quality controls are performed simultaneously with the PT test. The INRatio 2 is intended for use in non-sterile environments, such as patient homes or clinics, and is regulated under FDA 510(k) clearance (K072727) as a Class II medical device. Storage and handling follow manufacturer guidelines to ensure device integrity and accuracy.

Other devices from Hemosense, Inc.

  • Inratio Self-Test

    Class II
  • Inratio

    Class II

Questions

AI-generated answers, 20 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What clinical scenarios is the INRatio 2 PT Monitoring System primarily used for?

    The INRatio 2 is designed for real-time monitoring of prothrombin time (PT) and international normalized ratio (INR) in patients on anticoagulant therapy, particularly those using vitamin K antagonists. It is ideal for settings like outpatient clinics, anticoagulation management services, and home monitoring programs where quick, accurate results are needed to guide clinical decisions.

  • How is the INRatio 2 PT Monitoring System supplied, and is it reusable?

    The INRatio 2 is supplied as a single-use system, meaning each test cartridge and analyzer combination is intended for one use only. This design eliminates the need for cleaning or reprocessing, reducing the risk of contamination and ensuring consistent performance.

  • Does the INRatio 2 require calibration or additional control tests before use?

    No, the INRatio 2 does not require external calibration or separate control tests. Internal quality controls are performed simultaneously with the PT test, streamlining the workflow and ensuring reliable results without extra steps.

  • What type of environment is the INRatio 2 intended for, and how should it be stored?

    The INRatio 2 is designed for use in non-sterile environments, such as patient homes or clinics. Storage and handling should follow the manufacturer’s guidelines to maintain device integrity and accuracy, though specific conditions (e.g., temperature, humidity) are not detailed in the provided data.

  • What regulatory pathway was followed for the INRatio 2 PT Monitoring System, and what class is it assigned?

    The INRatio 2 was cleared through the FDA’s 510(k) pathway (K072727) as a Class II medical device, with a special clearance type and a decision of substantial equivalence. It falls under the Hematology specialty and is regulated under FDA regulation number 864.7750.