INRatio 2 PT Monitoring System
510(k) K072727
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK072727 | Date26 Oct 2007 | StatusActive |
Recalls
-
- Terminated Z-2357-2016
- Terminated Z-2360-2016
Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
-
- Terminated Z-0880-2015
- Terminated Z-0882-2015
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.
-
- Terminated Z-0188-2012
Alere has initiated a recall of the Professional Disposables International (PDI) alcohol prep products due to the potential presence of low levels of the microorganism Bacillus cereus.
-
- Terminated Z-1256-2009
INRatio Monitors that were manufactured with a specific EEPROM memory chip fail during use. The monitors report an error code and do not report a test result.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- GJS
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.7750
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 20 Sep 2026
The INRatio 2 PT Monitoring System is a point-of-care device designed for the measurement of prothrombin time (PT) and derived international normalized ratio (INR) values. It is classified as a blood coagulation test system, enabling rapid and accurate monitoring of patients on anticoagulant therapy, particularly those using vitamin K antagonists. The system provides real-time results, facilitating immediate clinical decision-making in settings such as outpatient clinics, anticoagulation management services, and home monitoring programs.
The device operates using a portable, handheld analyzer that requires a small blood sample applied to a test cartridge. Supplied as a single-use system, it does not require external calibration or additional control tests, as internal quality controls are performed simultaneously with the PT test. The INRatio 2 is intended for use in non-sterile environments, such as patient homes or clinics, and is regulated under FDA 510(k) clearance (K072727) as a Class II medical device. Storage and handling follow manufacturer guidelines to ensure device integrity and accuracy.
Other devices from Hemosense, Inc.
Questions
AI-generated answers, 20 Sep 2026
What clinical scenarios is the INRatio 2 PT Monitoring System primarily used for?
The INRatio 2 is designed for real-time monitoring of prothrombin time (PT) and international normalized ratio (INR) in patients on anticoagulant therapy, particularly those using vitamin K antagonists. It is ideal for settings like outpatient clinics, anticoagulation management services, and home monitoring programs where quick, accurate results are needed to guide clinical decisions.
How is the INRatio 2 PT Monitoring System supplied, and is it reusable?
The INRatio 2 is supplied as a single-use system, meaning each test cartridge and analyzer combination is intended for one use only. This design eliminates the need for cleaning or reprocessing, reducing the risk of contamination and ensuring consistent performance.
Does the INRatio 2 require calibration or additional control tests before use?
No, the INRatio 2 does not require external calibration or separate control tests. Internal quality controls are performed simultaneously with the PT test, streamlining the workflow and ensuring reliable results without extra steps.
What type of environment is the INRatio 2 intended for, and how should it be stored?
The INRatio 2 is designed for use in non-sterile environments, such as patient homes or clinics. Storage and handling should follow the manufacturer’s guidelines to maintain device integrity and accuracy, though specific conditions (e.g., temperature, humidity) are not detailed in the provided data.
What regulatory pathway was followed for the INRatio 2 PT Monitoring System, and what class is it assigned?
The INRatio 2 was cleared through the FDA’s 510(k) pathway (K072727) as a Class II medical device, with a special clearance type and a decision of substantial equivalence. It falls under the Hematology specialty and is regulated under FDA regulation number 864.7750.