Instatrak 3000
by Visualization Technology, Inc.
510(k) K983529
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK983529 | Date31 Dec 1998 | StatusActive |
Recalls
-
- Terminated Z-2251-2012
GE Healthcare Surgery had discovered that using the Inverted Headset Placement is not a validated configuration with the IT 3000, 2500, 2500 plus, 3500, 3500 plus, ConneCTstat, and ConneCTstat Plus Surgical Navigation equipment.
-
- Terminated Z-0709-06
Small retaining pin may detach and fall into the surgical field during stereotactic surgery.
-
- Terminated Z-0704-2007
Tracking inaccuracy of 5-10 mm may result when the shunt guide accessory is used with version 5.1 software.
-
- Terminated Z-0689-05
Enhanced sterility testing revealed the sterilization efficacy of accessories ('Sensors') used with certain Navigation and Visualization systems may be compromised.
-
- Terminated Z-0814-05
- Terminated Z-0815-05
Navigation inaccuracy may result if the headset registration method is used with the Axcess System Kit.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LLZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.2050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent