  "format": "devicerts.device.v1",
  "cite_as": "Iolmaster 500, Carl Zeiss Meditec AG. deviCERTS medical device catalogue. https://devicerts.com/devices/iolmaster-500-hjo (data as of 2026-11-25).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/iolmaster-500-hjo",
    "json_url": "https://devicerts.com/devices/iolmaster-500-hjo.json",
    "markdown_url": "https://devicerts.com/devices/iolmaster-500-hjo.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-25",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/iolmaster-500-hjo",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Iolmaster 500",
    "registry_name": "IOLMASTER 500",
    "url": "https://devicerts.com/devices/iolmaster-500-hjo",
    "manufacturer": {
      "name": "Carl Zeiss Meditec AG",
      "url": "https://devicerts.com/manufacturers/carl-zeiss-meditec-ag",
      "country": "Germany",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K101182",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HJO"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K101182",
      "number_label": null,
      "device_udi_di": null,
      "date": "2010-11-03",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0357-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-10-27",
      "reason": "IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.",
      "applies": "this device",
      "via": "K101182"
    },
    {
      "number": "Z-0358-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-10-27",
      "reason": "IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.",
      "applies": "this device",
      "via": "K101182"
    },
    {
      "number": "Z-0333-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-23",
      "reason": "The user database delivered for download contained an erroneous value. Specifically the \"aO\" constant of the lens AMO Sensar AR40E is listed as \"0.0\" in the internal lens database of the IOLMaster. The correct value is -2.420. This could lead to the implantation of lens with a wrong diopter value.",
      "applies": "this device",
      "via": "K101182"
    },
    {
      "number": "Z-0334-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-23",
      "reason": "The user database delivered for download contained an erroneous value. Specifically the \"aO\" constant of the lens AMO Sensar AR40E is listed as \"0.0\" in the internal lens database of the IOLMaster. The correct value is -2.420. This could lead to the implantation of lens with a wrong diopter value.",
      "applies": "this device",
      "via": "K101182"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K101182",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HJO",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 886.1850",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Ophthalmic",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}