Iolmaster
by Carl Zeiss, Inc. 510(k)
Active
510(k) K993357
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK993357 | Date20 Mar 2000 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0358-2016
- Terminated Z-0357-2016
IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
-
- Terminated Z-0333-2014
- Terminated Z-0334-2014
The user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is -2.420. This could lead to the implantation of lens with a wrong diopter value.
Specification
- 510(k) Number
- FDA Product Code
- HJO
- Device Class
- Class II
- Regulation
- 21 CFR Part 886.1850
- Advisory Committee
- Ophthalmic
- FDA Decision
- SESE Substantially Equivalent