Iq 200 System

by Intl. Remote Imaging Systems

510(k) K022774

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

10 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK022774 Date21 Oct 2002ClearedStatusActive

Recalls

14 recalls · 10 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-0160-2027 FDA Not yet classified initiated 12 Aug 2026 on shared 510(k) K022774
    • Ongoing Z-0161-2027 FDA Not yet classified initiated 12 Aug 2026 on shared 510(k) K022774

    The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.

  • 4 recalls, same reason

    • Ongoing Z-0784-2024 FDA Class II initiated 19 Dec 2023
    • Ongoing Z-0785-2024 FDA Class II initiated 19 Dec 2023
    • Ongoing Z-0786-2024 FDA Class II initiated 19 Dec 2023
    • Ongoing Z-0787-2024 FDA Class II initiated 19 Dec 2023

    Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.

  • 3 recalls, same reason

    • Ongoing Z-1599-2023 FDA Class II initiated 5 Apr 2023 on shared 510(k) K022774
    • Ongoing Z-1597-2023 FDA Class II initiated 5 Apr 2023 on shared 510(k) K022774
    • Ongoing Z-1598-2023 FDA Class II initiated 5 Apr 2023 on shared 510(k) K022774

    Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

    • Ongoing Z-1596-2023 FDA Class II initiated 5 Apr 2023 on shared 510(k) K022774

    Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.

    • Terminated Z-1945-2020 FDA Class II initiated 30 Mar 2020 on shared 510(k) K022774

    A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.

    • Terminated Z-0914-2020 FDA Class II initiated 28 Oct 2019

    Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).

    • Terminated Z-1362-2018 FDA Class II initiated 2 Mar 2018 on shared 510(k) K022774

    Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.

    • Terminated Z-1366-2018 FDA Class II initiated 10 Nov 2017 on shared 510(k) K022774

    iQ200 Series Urine Microscopy Analyzer may intermittently fail to read urine sample dilution barcode labels causing erroneous results that could delay treatment for health conditions such as hematuria.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LKM
Device Class
Class II
Regulation
21 CFR Part 864.5200
Advisory Committee
Hematology
FDA Decision
SESE Substantially Equivalent

Other devices from Intl. Remote Imaging Systems

See all devices from Intl. Remote Imaging Systems →