Skip to content

ISE Reagents, Glucose, CRP Latex, DxC 500 AU Clinical Chemistry Analyzer

by Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd.

510(k) K220977

Active European Union and United States

CE IVDR 510(k)

Valid until

Not listed

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

7 ongoing

Recall on shared 510(k) K220977; other models may be affected

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source EUDAMED, FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE 510(k) Substantially Equivalent · Traditional NumberK220977 Date20 Jul 2023ClearedStatusActive
European UnionCE/IVDR EUDAMED Number— Date—StatusActive

Recalls

7 recalls · 7 ongoing · FDA

    • Ongoing Z-0084-2027 FDA Class II initiated 2 Sep 2026 on shared 510(k) K220977

    Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.

    • Ongoing Z-2974-2026 FDA Class II initiated 15 Jul 2026 on shared 510(k) K220977

    a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

    • Ongoing Z-1784-2025 FDA Class II initiated 20 Mar 2025 on shared 510(k) K220977

    Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.

    • Ongoing Z-1672-2025 FDA Class II initiated 17 Mar 2025 on shared 510(k) K220977

    The reason for the recall is incorrect sample barcode identification and processing errors that result in the reporting of erroneous and delayed results.

    • Ongoing Z-1233-2025 FDA Class II initiated 10 Jan 2025 on shared 510(k) K220977

    Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.

    • Ongoing Z-2691-2024 FDA Class II initiated 10 Jul 2024 on shared 510(k) K220977

    Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.

    • Ongoing Z-2455-2024 FDA Class II initiated 5 Jun 2024 on shared 510(k) K220977

    There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. Issue 2: calibration with expired calibrator. They both may cause erroneous patient results, but the probability of serious adverse health consequences or death is unlikely.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/IVDR Active
  • United States FDA 510(k) Active

Specification

EUDAMED · 510(k)

510(k) Number
FDA Product Code
JGS
Device Class
Class II
Regulation
21 CFR Part 862.1665
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Overview

AI-generated summary, 18 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

This device is a set of ISE (Ion-Selective Electrode) reagents and latex agglutination reagents designed for use on the DxC 500 AU Clinical Chemistry Analyzer. The reagents include glucose and CRP (C-reactive protein) assays, enabling quantitative measurement of glucose levels and qualitative detection of CRP in serum or plasma samples. These reagents are intended for in vitro diagnostic use in clinical chemistry laboratories.

The reagents are supplied as single-use components for automated analysis on the DxC 500 AU analyzer, requiring no additional sterilization beyond manufacturer-provided packaging. They are classified as Class II devices under FDA regulations and comply with FDA 510K and IVDR requirements. Storage conditions and compatibility with the analyzer are specified in the manufacturer’s instructions, ensuring accurate and reliable diagnostic results.

Other devices from Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd.