Ivac Needle Free Valve Administration Sets
by Y 510(k)
Active
510(k) K960280
Regulatory status
Recall ongoing. 3 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK960280 | Date4 Apr 1996 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2460-2021
- Ongoing Z-2467-2021
- Ongoing Z-2466-2021
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
-
- Terminated Z-0221-2016
CareFusion is recalling the SmartSite Extension Set due to disconnection and leakage issues.
-
- Terminated Z-1053-2015
CareFusion is recalling the SmartSite Low Sorbing Infusion set due to disconnection and leakage issues.
-
- Terminated Z-1459-2014
Disconnections and leakages at the connection of the smallbore tubing and the SmartSite Y-port.
-
- Terminated Z-0925-2014
CareFusion is recalling the SmartSite Needlefree Connector, model number 2000E, because of connection issues. The affected SmartSite connector lots may unintentionally disconnect from a female luer, may be difficult to disconnect from a female luer, or may fail to disconnect from a female luer once attached. Leakage may also be observed if the connector disconnects from the female luer during in
Specification
- 510(k) Number
- FDA Product Code
- FPA
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5440
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent