  "format": "devicerts.device.v1",
  "cite_as": "Ivenix Infusion System 4.0.0, Fresenius Kabi. deviCERTS medical device catalogue. https://devicerts.com/devices/ivenix-infusion-system-ims-0700 (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/ivenix-infusion-system-ims-0700",
    "json_url": "https://devicerts.com/devices/ivenix-infusion-system-ims-0700.json",
    "markdown_url": "https://devicerts.com/devices/ivenix-infusion-system-ims-0700.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/ivenix-infusion-system-ims-0700",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Ivenix Infusion System 4.0.0",
    "registry_name": "Ivenix Infusion System 4.0.0",
    "url": "https://devicerts.com/devices/ivenix-infusion-system-ims-0700",
    "manufacturer": {
      "name": "Fresenius Kabi",
      "url": "https://devicerts.com/manufacturers/fresenius-kabi",
      "country": "Germany",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Catalogue number",
        "values": [
          {
            "value": "IMS-0700",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "00811505030351",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K183311",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "PHC",
      "FRN"
    ],
    "emdn": [],
    "gmdn": [
      {
        "code": "47903",
        "term": "Infusion pump medication-dosage application software"
      }
    ],
    "description": null,
    "noindex": true
  },
  "authorizations": [
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "00811505030351",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2022-09-07",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K183311",
      "number_label": null,
      "device_udi_di": null,
      "date": "2019-06-07",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2223-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-05-06",
      "reason": "Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-2224-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-05-06",
      "reason": "Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-2274-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-02-04",
      "reason": "Potential for the Cassette Loading Lever to break.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-1048-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-11-21",
      "reason": "Emphasizing instructions for LVP duration programming located in the IFU.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-0885-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-11-14",
      "reason": "Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-0586-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-11-03",
      "reason": "Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-0848-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-11-03",
      "reason": "Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-0378-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-09-03",
      "reason": "Potential issue that can cause the device to register \"phantom\" touches in the lower-left corner of the touchscreen during operation.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-2232-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-06-24",
      "reason": "Potential for external cassette leaks",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-2233-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-06-24",
      "reason": "Potential for external cassette leaks",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-2234-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-06-24",
      "reason": "Potential for external cassette leaks",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-2235-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-06-24",
      "reason": "Potential for external cassette leaks",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-1869-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-05-12",
      "reason": "Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-1019-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-10",
      "reason": "Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: \n1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. \n2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-0763-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-12-06",
      "reason": "Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-3153-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-08-30",
      "reason": "The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy.  These could cause serious patient harm or death.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-2721-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-08-01",
      "reason": "A defect may cause an uncontrolled medication flow condition which may result in a dose greater than prescribed for the patient (i.e., medication overdose).",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-1484-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-03-07",
      "reason": "Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-0549-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-11-29",
      "reason": "The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-1283-2024",
      "classification": "Class II",
      "status": "completed",
      "initiated": "2023-05-10",
      "reason": "Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to delay in therapy. 2) Alert is not annunciated informing the clinician that the bolus cannot be delivered when the entered bolus dose exceeds the Care Profile Hard Rate Max limit and Rapid Bolus is selected. May lead to over infusion. Both issues were resolved in all fielded product in software version 5.8.0, which was installed in affected units May thru August 2023.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-1313-2023",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-03-10",
      "reason": "Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-1282-2024",
      "classification": "Class II",
      "status": "completed",
      "initiated": "2022-09-16",
      "reason": "Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm.\n\nThis issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.",
      "applies": "shared submission",
      "via": "K183311"
    },
    {
      "number": "Z-0002-2023",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2022-09-08",
      "reason": "The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.",
      "applies": "this device",
      "via": "K183311"
    },
    {
      "number": "Z-1381-2022",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2022-04-22",
      "reason": "Downstream occlusion alarm is sometimes immediately followed by a pump problem alarm.",
      "applies": "this device",
      "via": "K183311"
    },
    {
      "number": "Z-2635-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-04-22",
      "reason": "Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set",
      "applies": "shared submission",
      "via": "K183311"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "Ivenix Infusion System",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Infusion pump medication-dosage application software",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / Ref / Catalog",
          "values": [
            {
              "value": "4.0.0",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "IMS-0700",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "00811505030351",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K183311",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "PHC",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Infusion pump medication-dosage application software",
              "code": "47903",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "880.5725",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}