  "format": "devicerts.device.v1",
  "cite_as": "Ivue, Model 100, Optovue, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/ivue-model-100-obo (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/ivue-model-100-obo",
    "json_url": "https://devicerts.com/devices/ivue-model-100-obo.json",
    "markdown_url": "https://devicerts.com/devices/ivue-model-100-obo.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/ivue-model-100-obo",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Ivue, Model 100",
    "registry_name": "IVUE, MODEL 100",
    "url": "https://devicerts.com/devices/ivue-model-100-obo",
    "manufacturer": {
      "name": "Optovue, Inc.",
      "url": "https://devicerts.com/manufacturers/optovue-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K091404",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "OBO"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K091404",
      "number_label": null,
      "device_udi_di": null,
      "date": "2010-04-22",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0024-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-09-07",
      "reason": "Devices labeled for an intended use not included in the existing 510(k) or PMA. Specifically, \"The Retina\" poster provided with the systems, and the book \"RETINAL OCT Analysis and Interpretation Method\" by Bruno Lumbroso and Marco Rispoli which are not labeled within 51O(k) cleared product description.",
      "applies": "this device",
      "via": "K091404"
    },
    {
      "number": "Z-1004-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2010-07-20",
      "reason": "Use of the CAM attachment may result in a positioning anomaly and mistreatment in the eyes of the patient.",
      "applies": "shared submission",
      "via": "K091404"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K091404",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "OBO",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 886.1570",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Ophthalmic",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}