Jostra Hl-20 Integrated Perfusion System
510(k) K943803
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK943803 | Date30 Jan 1996 | StatusActive |
Recalls
-
- Terminated Z-3057-2011
HL 20 Consoles for 4-pump and 5-pump configuration can malfunction if an excessive amount of liquid is used causing circuit board malfunction.
-
- Terminated Z-3084-2011
Maquet HL 20 Twin Pump Modules may unintentionally increase speeds for a 1.5-2 second period of time before returning to its original speed.
-
- Terminated Z-1394-2011
A wrong ball bearing was installed in the pump heads of product. This may make the pump head wear out and cease function.
-
- Terminated Z-0159-2007
The HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in the tube guide rolls, they cannot roll smoothly and finally they can get stuck.
-
- Terminated Z-0158-2007
The hand crank necessary for the emergency manual operation of the pump is too wide diametrically to fit into the holes which they were designed to fill.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DTQ
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4220
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent