Skip to content

KARL STORZ UDEL Sterilization Trays

by Karl Storz Endoscopy-America, Inc.

510(k) K201945

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK201945 Date25 Feb 2021ClearedStatusActive

Recalls

1 recall · 1 ongoing · FDA

    • Ongoing Z-0442-2022 FDA Class II initiated 15 Oct 2021

    To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximum length of 845mm and 2) ethylene oxide is excluded as a sterilization modality.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
KCT
FDA applicant
Karl Storz Endoscopy America, Inc.
Device Class
Class II
Regulation
21 CFR Part 880.6850
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent