Kinetdx Picture Archiving and Communications System

by Acuson Corp.

510(k) K023772

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

5 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK023772 Date22 Nov 2002ClearedStatusActive

Recalls

5 recalls · FDA

    • Terminated Z-1601-2016 FDA Class II initiated 14 Apr 2016

    Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.

    • Terminated Z-0826-2016 FDA Class II initiated 29 Jan 2016

    Siemens is releasing a software update that addresses an issue of mixing data from multiple patients. In rare situations, echo trend graphs may mix data from multiple patients.

    • Terminated Z-1183-2016 FDA Class II initiated 25 Jan 2016

    Siemens' conducting a recall due to a potential issue when using the measurement package of the VA10 version of syngo Dynamics.

    • Terminated Z-1613-2014 FDA Class II initiated 3 Apr 2014

    Systems with the affected serial number are experiencing a software error which may lead to incorrect patient demographics display.

    • Terminated Z-1216-2014 FDA Class II initiated 27 Jan 2014

    Potential patient data mixup. Improper error handling could allow for two different patients to be displayed in syngo Dynamics. Images shown for one patient and reporting system shown for a different patient, leading to the potential for misdiagnosis.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LLZ
Device Class
Class II
Regulation
21 CFR Part 892.2050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Acuson Corp.

See all devices from Acuson Corp. →