Kodak Directview Dr 3000 System 64KW/80KW Generators, Models 6551345, 6551386
510(k) K060550
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK060550 | Date31 Mar 2006 | StatusActive |
Recalls
-
- Terminated Z-0806-2020
Carestream Health has discovered a potential safety problem that can lead to unintended movement of the U-Arm.
-
- Terminated Z-0731-2008
Unintended Positioner Movement - The U-Arm Positioner starts to move 1) without command from the operator and/or 2) can unexpectedly start movement if the system is powered down and re-energized after a collision has occurred with the Quantum Table.
-
- Terminated Z-0717-2007
The x-ray systems do not indicate the field size when the system's detector is turned off and the collimator is in manual mode. This results in the systems being noncompliant with the field indication performance standards of section 1020.31(e)(2)& (3) of 21Code of Federal Regulations (CFR).
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KPR
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1680
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent