Skip to content

Kodak Ektachem 250 Analyer

by Eastman Kodak Company

510(k) K922072

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

15 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK922072 Date10 Jul 1992ClearedStatusActive

Recalls

15 recalls · FDA

  • 4 recalls, same reason

    • Terminated Z-2174-2019 FDA Class II initiated 8 Jul 2019
    • Terminated Z-2176-2019 FDA Class II initiated 8 Jul 2019
    • Terminated Z-2175-2019 FDA Class II initiated 8 Jul 2019
    • Terminated Z-2177-2019 FDA Class II initiated 8 Jul 2019

    Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System

  • 4 recalls, same reason

    • Terminated Z-1935-2019 FDA Class II initiated 8 Apr 2019
    • Terminated Z-1936-2019 FDA Class II initiated 8 Apr 2019
    • Terminated Z-1933-2019 FDA Class II initiated 8 Apr 2019
    • Terminated Z-1934-2019 FDA Class II initiated 8 Apr 2019

    Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.

  • 2 recalls, same reason

    • Terminated Z-1072-2016 FDA Class II initiated 21 Dec 2015
    • Terminated Z-1071-2016 FDA Class II initiated 21 Dec 2015

    Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).

    • Terminated Z-1226-2014 FDA Class II initiated 9 Jan 2014 on shared 510(k) K922072

    An incorrect assay value (0.3785) was printed on the product carton for the 460 nm wavelength (filter) that may cause a delay in the ability to obtain and report TBIL, Bu, Bc, and derived tests results. The correct assay value is 0.8658.

    • Terminated Z-1088-2014 FDA Class II initiated 17 Oct 2013 on shared 510(k) K922072

    Ortho Clinical Diagnostics has recalled VITROS Chemistry Products Na+ Slides and the Calibrator Kit 2 due to false positive serum and plasma sample results.

  • 2 recalls, same reason

    • Terminated Z-0296-2014 FDA Class II initiated 12 Jun 2013 on shared 510(k) K922072
    • Terminated Z-0305-2014 FDA Class II initiated 12 Jun 2013 on shared 510(k) K922072

    Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.

    • Terminated Z-1277-2012 FDA Not yet classified initiated 2 Feb 2012 on shared 510(k) K922072

    Ortho Clinical Diagnostics received complaints of lower than expected results (< 6 U/L) for samples that were known to be outside of the measuring (reportable) range (> 1000 U/L) when using Calibration Diskette/Assay Data Diskette Data Release Version (DRV) 5653 through 5666 and VITROS Chemistry Products ALT Slides, Generation (GEN) 26 (Product Code 1655281).

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JJE
Device Class
Class I
Regulation
21 CFR Part 862.2160
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Other devices from Eastman Kodak Company