Kodak Ektachem Clinical Chemistry Slide Phenytoin
510(k) K941142
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK941142 | Date13 Jun 1994 | StatusActive |
Recalls
-
- Terminated Z-2344-2018
There is a potential for imprecise results when using the slides. Increase in imprecision when using PHYT coating 0165. This leads to higher and lower than expected results for both quality control samples and patient samples.
-
- Terminated Z-0273-2018
The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.
-
- Terminated Z-0202-2016
The firm is conducting a product correction due to an increase in U90-382 or 6LU condition codes (wash errors) when using VITROS Calibrator Kit 9, Lot 0954.
-
- Terminated Z-0007-2015
Ortho Clinical Diagnostics identified a potential for biased results to be generated when using VITROS PHYT Slides, Lot 2613-0150-4913. Some positively biased results were identified during testing of a different VITROS PHYT Slide lot; the affected slides were not released for distribution. Since Lot 2613-0150-4913 was manufactured in the same event, there is the potential for slides from this lot to also be affected.
-
- Terminated Z-0321-2014
Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DIP
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.3350
- Advisory Committee
- Clinical Toxicology
- FDA Decision
- SESE Substantially Equivalent