L10-5v Transducer 11319682
by Siemens Medical Solutions USA, Inc.
Ref 11319682Model MDR_07_IIA_IQH_BTS LB A_ACDV_L10-5v_11319682
L10-5v Transducer 11319682 is an extracorporeal ultrasound imaging transducer (EU MDR Class IIa; FDA Class II; FDA product code ITX, Transducer, ultrasonic, diagnostic) made by Siemens Medical Solutions USA, Inc., registered in EUDAMED on 8 Sep 2026 (Basic UDI-DI 0405686902392VX), cleared under FDA 510(k) K233610 on 12 Jun 2024 (2 earlier submissions) and listed in FDA GUDID since 10 Mar 2018 (UDI-DI 04056869055466), active as of 23 Oct 2026.
Active
Valid until
No date given
FDA recalls
2 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0405686902392VX 04056869055466 | Date8 Sep 2026 | StatusActive |
| United States | FDA 510(k)SESE GUDID | NumberK233610 | Date12 Jun 2024 | StatusActive |
| United States | GUDID listing GUDID | Number04056869055466 | Date10 Mar 2018 | StatusListed |
Official records
- EUDAMED public site 0405686902392VX
- AccessGUDID record 04056869055466
- FDA 510(k) database K233610
- FDA 510(k) database K232617
- FDA 510(k) database K163396
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI 0405686902392VX registered 8 Sep 2026 Active
- FDA 510(k) K233610 decided 12 Jun 2024 Active
- FDA GUDID UDI-DI 04056869055466 listed 10 Mar 2018 Listed
Recalls
-
- Ongoing Z-3217-2024
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
-
- Ongoing Z-1601-2024
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
Timeline
- Cleared FDA 510(k) K163396
- Listed FDA GUDID 04056869055466
- Cleared FDA 510(k) K232617
- Recall initiated FDA recalls Z-1601-2024
- Cleared FDA 510(k) K233610
- Recall initiated FDA recalls Z-3217-2024
- Registered EUDAMED 0405686902392VX
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- L10-5v Transducer
- Generic name
- Extracorporeal ultrasound imaging transducer, hand-held
- Device name
- MDR_07_IIA_IQH_BTS LB A_ACDV_L10-5v_11319682
- Model / reference
- 11319682, MDR_07_IIA_IQH_BTS LB A_ACDV_L10-5v_11319682
- Primary DI / UDI-DI
- 04056869055466
- Basic UDI-DI
- 0405686902392VX
- FDA product code
- ITX
- EMDN code
- Z110402010101 Two–dimensional linear–array transcutaneous ultrasound probes
- GMDN term
- Extracorporeal ultrasound imaging transducer, hand-held
- Device class
- Class IIa EUDAMEDClass II GUDID
- Regulation number
- 892.1570
- EU registration type
- Standard device
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IIa | Class II |
Similar devices in this category
Other devices filed under TWO–DIMENSIONAL LINEAR–ARRAY TRANSCUTANEOUS ULTRASOUND PROBES.
- HFL50xp/15-6 MHz P14567-22A by Fujifilm Sonosite, Inc.
- HFL38x/13-6 MHz P07682-70A by Fujifilm Sonosite, Inc.
- Linear Probe L14-4 A8J9 by KONICA MINOLTA, INC.
- SP-L01 99-7901-00 by Interson Corporation
- SR 7.5 MHz 99-5554-00 by Interson Corporation
- L4-20t-D 5874154 by GE Medical Systems CO. Ltd
- Vscan Air SL by Ge Vingmed Ultrasound
- L64 04573596212172 by FUJIFILM CORPORATION
- Linear Probe L11-3 A9E5 by KONICA MINOLTA, INC.
- LP 4-13 120000026 by Esaote S.p.A.