L52H
Ref 04547410566949
Active
Valid until
Not listed
FDA recalls
3 ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number454741050100000000000069Y 04547410566949 | Date18 Feb 2026 | StatusActive |
| United States | GUDID listing GUDID | Number04547410566949 | Date29 Oct 2025 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-0190-2022
- Ongoing Z-0189-2022
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
-
- Ongoing Z-2308-2021
The screw cover at the tip of the probe may fall off after a cleaning and sterilization process. Users advised to conduct a preventive inspection of the probe before and after sterilization/disinfection/cleaning and before and after an interoperative procedure.
-
- Terminated Z-1282-2021
Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Generic Name
- Extracorporeal ultrasound imaging transducer, hand-held
- Model / Reference
- 04547410566949 EUDAMEDL52H GUDID
- Primary DI / UDI-DI
- 04547410566949
- Basic UDI-DI
- 454741050100000000000069Y
- Direct Marketing DI
- 04547410566949
- FDA Product Code
- IYO, IYN, ITX
- EMDN Code
- Z110402010101 Two–dimensional linear–array transcutaneous ultrasound probes
- GMDN Term
- Extracorporeal ultrasound imaging transducer, hand-held
- Device Class
- Class IIa EUDAMEDClass II GUDID
- Regulation Number
- 892.1560, 892.1550, 892.1570
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
- Sterilization Method
- High-level Disinfectant
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | 04547410566949 | L52H |
| Device Class | Class IIa | Class II |
Manufacturer details
- EUDAMED SRN
- JP-MF-000010401
- Authorised representative
- FUJIFILM Healthcare Europe GmbH DE-AR-000040920
Other devices from Fujifilm Corporation
Properties
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No