Laryngeal Surface Electrode-Endotracheal Tube, M-500-Et, Lse-Et
510(k) K003745
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK003745 | Date16 Jan 2001 | StatusActive |
Recalls
-
- Terminated Z-0798-2010
This recall has been initiated due to a conductive glue used to attach the lead wire to electrode, which has been found in a small number of electrodes, to cause a short circuit in one of the channels. This is due to migration of the adhesive from one pad to the other creating an electrical short. Use of one of these defective electrodes could lead to a failure to monitor. Our investigation has c
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- ETN
- Device Class
- Class II
- Regulation
- 21 CFR Part 874.1820
- Advisory Committee
- Ear, Nose, Throat
- FDA Decision
- SESE Substantially Equivalent