Lci Option for Innocor
510(k) K102047
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK102047 | Date1 Nov 2011 | StatusActive |
Recalls
-
- Terminated Z-2769-2016
The purpose of the recall is to replace data transmission cable (USB-cable) connecting the lnnocor device with an external computer. The replacement USB cable will provide galvanic isolation of the lnnocor from the external computer via opto-couplers, as opposed to the grounding protection currently in use. The replacement is intended to prevent risk of electric shock.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- BZC
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.1880
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent