LeadCare Plus Blood Lead Testing System
by Magellan Diagnostics United States 510(k)
510(k) K142705
Product code DOF
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK142705 | Date7 Jul 2015 | StatusActive |
Coverage
- Active
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1894-2026
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
-
- Terminated Z-1953-2021
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
-
- Terminated Z-1296-2018
The incorrect placement of AA batteries into certain LeadCare Plus Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
-
- Terminated Z-1926-2017
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Specification
- 510(k) Number
- FDA Product Code
- DOF
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.3550
- Advisory Committee
- Clinical Toxicology
- FDA Decision
- SESE Substantially Equivalent